ASTM ER Standard Test Method for Determination of Hydrogen in Titanium and Titanium Al_材料科学_工程科技_专业资料。. ASTM E() Standard Test Method for Determination of Hydrogen in Titanium and Titanium Alloys by Inert Gas Fusion Thermal Conductivity/Infrared. Buy ASTM E TEST METHOD FOR DETERMINATION OF HYDROGEN IN TITANIUM AND TITANIUM ALLOYS BY INERT GAS.

Author: Kat Mikakasa
Country: Spain
Language: English (Spanish)
Genre: Travel
Published (Last): 21 September 2016
Pages: 30
PDF File Size: 10.91 Mb
ePub File Size: 9.82 Mb
ISBN: 159-4-48258-631-8
Downloads: 18529
Price: Free* [*Free Regsitration Required]
Uploader: Nalkis

Chemical Testing

In this case instrument response must always be veri? Precision and Bias This titanium hydrogen reference material should have a hydrogen content greater than or approximately equal to the unknown samples within the scope of this test method 0.

The mid-range material must have a hydrogen concentration that is above the limit of zstm, but below that of the high calibration point, preferably as close to the mid-point of the calibration curve as possible. This cannot be corrected by making adjustments to the instrument.

Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk of infringement of such rights, are entirely their own responsibility. Specimens must be of an appropriate size to? Treat each specimen as directed in Standard Test Method f The data are presented in Table 1. NOTE 3—Regardless of the method used, the sample must not be allowed to overheat, as this will adversely affect the results of the analysis.

  CAKES AND ALE SOMERSET MAUGHAM PDF

Chemical Testing at WMT&R Inc.

Use a fresh crucible each time. No other units of measurement are included in this standard. ASTM D D Standard Reference Material: The accuracy of this test method is largely dependent upon the absence of bias in the hydrogen values assigned to the reference materials and upon the homogeneity of these materials.

Six instruments were of the thermal conductivity hydrogen measurement system con? Current edition approved Aug. In the absence of bias among the reference materials, the average result for this reference material should agree with the certi? It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

For specimens of unknown 2 The ILS used an acceptance criterion where the value fell within the allowable limits of the certi? If you feel that your comments have not received a fair hearing you should make your views known to the ASTM Committee on Standards, at the address shown below.

The acceptance range for the veri? NOTE 14—Typically, repeatability standard deviation is a function of the concentration of the analyte. Your comments will receive careful consideration at a meeting of the responsible technical committee, which you may attend.

  L6561 APPLICATION NOTE PDF

The user should determine and correct the cause, and repeat Dial the sample weight or a multiple of the sample weight on the weight compensator and use the where: Make a minimum of two determinations using a specimen as directed in NOTE 15—The presence of bias between the reference material used in The prediction interval is de?

Results must be supported by acceptable preceding and subsequent veri? It is assumed that all who use this test method will be trained analysts capable of performing common laboratory procedures skillfully and safely. Originally approved in The testing and statistical analysis were performed in accordance with the provisions of Practice E, Plan A. ASTM International takes no aatm respecting the validity of any patent rights asserted in connection with any item mentioned in e14447 standard.

Your comments are invited either for revision of this standard or for additional standards and should be addressed to ASTM International Headquarters.